TST90 iTHERM ModuLine Industrial Modular Thermometer

$540.00

Endress+Hauser

TST90 is mainly used in the chemical industry but also finds its use in other branches. The device is used for high-precision measurement of temperature
differences with matched sensor pairs. A variety of dimensions offers flexible application possibilities.

  • Universally usable for thermal energy measurements

  • Measuring range: -50 to +400 °C (-58 to +752 °F)

 

Description

Field of application

The thermometer is mainly used in the chemical industry but also finds its use in other branches. The device is used for high-precision measurement of temperature differences with matched sensor pairs. A variety of dimensions offers flexible application possibilities.

  • Universally usable for thermal energy measurements

  • Measuring range: -50 to +400 °C (-58 to +752 °F)

  • Paired in the range: 0 to +120 °C (+32 to +248 °F), Deviation of the sensors: ± 0.05 K

  • Paired in the range: –40 to 0 °C (-40 to 32 °F), Deviation of the sensors: ± 0.1 K

  • Protection class: up to IP68

Benefits

  • Fast replaceability of the insert

  • High degree of insert compatibility and design as per DIN 43772

  • Fast response time with reduced tip

Additional information

Brand

Endress+Hauser/E+H

Measuring principle

Resistance Temperature Detector

Characteristic / Application

metric style
modular temperature assembly (paired)
threaded process connection
with neck
incl. thermowell / protection tube (metal)

Thermowell / protection tube

welded protection tube

Insert / probe

mineral insulated (MI), flexible

Outer diameter protection tube / Insert

9,0 mm (0,35'')

Max. immersion length on request

up to 4.000,00 mm (157,48'')

Material protection tube/ thermowell

1.4571 (316Ti)

Optional coating

Not defined

Process connection

male thread:
G1/2"

Tip shape

straight
tapered

Surface roughness Ra

Not defined

Operating temperature range

PT 100:
-50 °C …400 °C
(-58 °F …752 °F)

Max. process pressure (static)

at 20 °C: 50 bar (725 psi)

Accuracy

+/- 0.05 K (refering to delta-T)

Response time

depending on configuration
t50 = 10 s
t90 = 31 s

Integration head transmitter

no