TH12 iTHERM CableLine Cable Probe

$200.00

Endress+Hauser

The TH12 temperature sensors are RTD’s designed for use in the process industry or factory automation type of environment. They are made up of a measurement probe with a RTD, insulation and sheath.

Among other applications the sensors can be used on:

  • Heat exchangers

  • Power & recovery areas

  • Carbon regeneration furnaces

  • Dryers

  • Flue gas

  • Compressor stations

  • Process reactors

 

SKU: TH12 Categories: , Tags: ,

Description

Field of application

The Pt100 cable probe is easy to install and provide a high operational safety due to reliable and accurate temperature measurement in common processes. Without additional thermowell the probe sheath is directly in contact with the process medium. This enables the cable probe to detect rapid temperature changes fast and accurate.

The TH12 temperature sensors are RTD’s designed for use in the process industry or factory automation type of environment. They are made up of a measurement probe with a RTD, insulation and sheath.

Among other applications the sensors can be used on:

  • Heat exchangers

  • Power & recovery areas

  • Carbon regeneration furnaces

  • Dryers

  • Flue gas

  • Compressor stations

  • Process reactors

Benefits

  • High flexibility due to customized immersion length

  • Fast response time with reduced/tapered tip form

  • Simplified model structure: Competitively priced, offers great value. Easy to order and reorder

  • Improved galvanic isolation on most devices (2 kV)

Additional information

Brand

Endress+Hauser/E+H

Measuring principle

Resistance Temperature Detector

Characteristic / Application

US style
cable probe
process connection as compression fitting
without neck

Thermowell / protection tube

without (not intended to use with thermowell)

Insert / probe

mineral insulated (MI), flexible

Outer diameter protection tube / Insert

1/8'' (3,18 mm)
3/16'' (4,76 mm)
1/4'' (6,35 mm)
3/8'' (9,53 mm)

Max. immersion length on request

up to 96" (2439 mm)
others on request

Material protection tube/ thermowell

Sensor sheath
316/316L

Process connection

compression fitting:
NPT1/8"
NPT1/4"
compression fitting:
G1/8"
G1/4"
G1/2"
NPT1/4"
NPT1/2"
M10x1
M8x1

Tip shape

straight
reduced

Operating temperature range

PT100 WW:
-200 °C … 600 °C
(-328 °F … 1.112 °F)
PT100 TF:
-50 °C … 200 °C
(-58 °F … 392 °F)

Max. process pressure (static)

at 20 °C: 40 bar (580 psi)
depends on configuration

Accuracy

class A acc. to IEC 60751
class B acc. to IEC 60751

Response time

63% rt = 2,0 s

Integration head transmitter

no

Reviews

There are no reviews yet.

Be the first to review “TH12 iTHERM CableLine Cable Probe”

Your email address will not be published. Required fields are marked *