TC65 may be used direct contact without thermowell or where the application requires it, spring load inside of a separate thermowell. Flexible configuration possible.

  • Fine Chemicals

  • Petrochemical industry

  • Power stations

  • Environmental engineering

 

Description

Field of application

The robust thermometer is designed for use in demanding and safety relevant applications e.g. in the Chemical, Oil & Gas and Energy industry. An optional head transmitter with all common communication protocols makes the device ready to use with enhanced measurement accuracy and reliability compared to directly wired sensors. The sensor may be used direct contact without thermowell or where the application requires it, spring load inside of a separate thermowell. Flexible configuration possible.

  • Fine Chemicals

  • Petrochemical industry

  • Power stations

  • Environmental engineering

  • Measuring range: –40 to 1 100 °C (–40 to 2 012 °F)

  • Static pressure range up to 75 bar dependent on the used process connection

  • Protection class: IP68

Benefits

  • Types of protection for use in hazardous locations: Intrinsic safety (Ex ia), flameproof (Ex d), non-sparking (Ex nA)

  • High degree of insert compatibility and design as per DIN 43772

  • High degree of flexibility thanks to modular design with standard terminal heads as per DIN EN 50446 and customer-specific immersion lengths

  • For plug-in/screw-in with sliding compression fitting

Additional information

Brand

Endress+Hauser/E+H

Measuring principle

Thermocouple

Characteristic / Application

metric style
modular temperature assembly
for heavy duty applications
suitable for hazardous areas
threaded process connection
without neck

Thermowell / protection tube

without (not intended to use with thermowell)

Insert / probe

mineral insulated (MI), flexible

Outer diameter protection tube / Insert

3,0 mm (0,12'')
6,0 mm (0,24'')

Max. immersion length on request

up to 30.000,00 mm (1.181,10'')

Material protection tube/ thermowell

1.4404 (316L)
Alloy 600 (2.4816)

Process connection

male thread:
NPT1/2"
NPT3/4"
compression fitting:
NPT1/2"
NPT3/4"

Tip shape

straight
tapered

Operating temperature range

Type K:
max. 1.100 °C
(max. 2.012 °F)
Type J:
max. 800 °C
(max. 1.472 °F)

Max. process pressure (static)

at 20 °C: 80 bar (1,160 psi)

Accuracy

class 1 acc. to IEC 60584

Response time

depending on configuration
t50 = 3 s
t90 = 7 s

Integration head transmitter

yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)

Ex - approvals

ATEX II
IECEx
NEPSI
EAC Ex
Explosion proof

Certification

Gost Metrology
SIL (transmitter only)

Reviews