Brand |
Endress+Hauser/E+H
|
Measuring principle |
Ultrasonic flow
|
Product Headline |
Highly robust gas specialist for fluctuating process conditions as remote version with up to 4 I/Os.
Flexible device with user-definable gas mixtures for demanding measuring tasks.
Accurate measurement of natural and process gas in the chemical as well as oil and gas industries.
|
Sensor features |
Maximum reliability even with humid or wet gas – sensor design insensitive to condensate. High-performance process control – real-time pressure- and temperature-compensated values. Efficient solution – multivariable, no pressure loss.
Direct measurement: flow, pressure & temperature. Wetted parts: titanium / 316L. Maximum measuring accuracy: 0.5 %.
|
Transmitter features |
Full access to process and diagnostic information – numerous, freely combinable I/Os. Reduced complexity and variety – freely configurable I/O functionality. Integrated verification – Heartbeat Technology.
Remote version with up to 4 I/Os. Backlit display with touch control and WLAN access. Standard cable between sensor and transmitter.
|
Nominal diameter range |
DN 25 to 300 (1 to 12")
|
Wetted materials |
Measuring tube: 1.4408/1.4409 (CF3M)
Transducer: 1.4404 (316, 316L, Titan Grade 2
|
Measured variables |
Volume flow, corrected volume flow, mass flow, flow velocity, speed of sound, pressure, temperature, density, dynamic viscosity, energy flow, Wobbe index, methane fraction, calorific value, molar mass
|
Max. measurement error |
Volume flow (standard):
‐ ±1.0 % o.r. for 3 to 40 m/s (9.84 to 131.23 ft/s)
‐ ±2 % o.r. for 0.3 to 3 m/s (0.98 to 9.84 ft/s)
Volume flow (optional calibration):
‐ ±0.5 % o.r. for 3 to 40 m/s (9.84 to 131.23 ft/s)
‐ ±1.0 % o.r. for 0.3 to 3 m/s (0.98 to 9.84 ft/s)
Corrected volume flow (standard):
‐ ±1.5 % o.r. for 3 to 40 m/s (9.84 to 131.23 ft/s)
‐ ±2.5 % o.r. for 0.3 to 3 m/s (0.98 to 9.84 ft/s)
Corrected volume flow (optional calibration):
‐ ±1.0 % o.r. for 3 to 40 m/s (9.84 to 131.23 ft/s)
‐ ±1.5 % o.r. for 0.3 to 3 m/s (0.98 to 9.84 ft/s)
Sound Velocity: ±0.2 % o.r.
|
Measuring range |
Gas: 0.3 m/s to 40 m/s
|
Max. process pressure |
0.7 to 101 bar a (10.15 to 1464.88 psi a)
|
Medium temperature range |
-50 to 150 °C (-58 to +302°F)
-50 to 100 °C (-58 to +212°F) with integrated pressure cell
|
Ambient temperature range |
-40 to 60 °C(-40 to +140 °F)
Optional: -50 to 60 °C(-58 to +140 °F)
|
Sensor housing material |
Stainless Steel, 1.4404(316/316L), 1.4408/1.4409 (CF3M)
|
Transmitter housing material |
AlSi10Mg, coated; 1.4409 (CF3M) similar to 316L
Polycarbonate
|
Degree of protection |
Sensor remote version: IP66/67, type 4X enclosure
Transmitter remote version: IP66/67, Type 4X enclosure
|
Display/Operation |
4-line backlit display with Touch Control (operation from outside)
Configuration via local display and operating tools possible
|
Outputs |
4 outputs:
4-20 mA HART (active/passive)
4-20 mA (active/passive)
Pulse/frequency/switch output (active/passive)
Double pulse output (active/passive)
Relay output
|
Inputs |
Status input
4-20 mA input
|
Digital communication |
HART, Modbus RS485
|
Power supply |
AC 100 to 230 V / DC 24 V (non-hazardous area)
|
Hazardous area approvals |
ATEX, IECEx, cCSAus, JPN, EAC, UK Ex, KC
|
Product safety |
CE, C-tick
|
Functional safety |
Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
|
Metrological approvals and certificates |
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a
|
Pressure approvals and certificates |
PED, CRN
|
Material certificates |
3.1 material
NACE MR0175/MR0103
|
Reviews