Brand |
Endress+Hauser/E+H
|
Measuring principle |
Coriolis
|
Product Headline |
Robust flowmeter with genuine loop-powered technology.
Highest measurement performance for liquids and gases in a wide range of applications.
|
Sensor features |
Highest process safety – immune to fluctuating and harsh environments. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in-/outlet run needs.
Mass flow: measured error ±0.1 %. Medium temperature up to +205 °C (+401 °F). Nominal diameter: DN 8 to 80 (³⁄₈ to 3").
|
Transmitter features |
Convenient device wiring – separate connection compartment. Safe operation – no need to open the device due to display with touch control, background lighting. Integrated verification – Heartbeat Technology.
Loop-powered technology. Robust dual-compartment housing. Plant safety: worldwide approvals (SIL, Haz. area).
|
Nominal diameter range |
DN 8 to 80 (⅜ to 3")
|
Wetted materials |
Measuring tube: 1.4539 (904L); 1.4404; Alloy C22, 2.4602 (UNS N06022)
Connection: 1.4404 (316/316L); Alloy C22, 2.4602 (UNS N06022)
|
Measured variables |
Mass flow, density, temperature, volume flow, corrected volume flow, reference density
|
Max. measurement error |
Mass flow (liquid): ±0.1 %
Volume flow (liquid): ±0.1 %
Mass flow (gas): ±0.25 % Density (liquid): ±0.0005 g/cm³
|
Measuring range |
0 to 70 000 kg/h (0 to 2570 lb/min)
|
Max. process pressure |
PN 100, Class 600, 63K
|
Medium temperature range |
Standard: –50 to +150 °C (–58 to +302 °F)
Option: –50 to +205 °C (–58 to +401 °F)
|
Ambient temperature range |
–40 to +60 °C (–40 to +140 °F)
|
Sensor housing material |
1.4301/1.4307 (304L), corrosion resistant
|
Transmitter housing material |
AlSi10Mg, coated, 1.4404 (316L)
|
Degree of protection |
IP66/67, type 4X enclosure
|
Display/Operation |
4‐line backlit display with touch control
(operation from outside)
Configuration via local display and operating tools possible
Remote display available
|
Outputs |
4‐20 mA HART (passive)
4‐20 mA (passive)
Pulse/frequency/switch output (passive)
|
Inputs |
None
|
Digital communication |
HART, PROFIBUS PA, FOUNDATION Fieldbus
|
Power supply |
DC 18 to 35 V (4‐20 mA HART with/without
pulse/frequency/switch)
DC 18 to 30 V (20 mA HART, 4‐20 mA)
DC 9 to 32 V (PROFIBUS PA)
|
Hazardous area approvals |
ATEX, IECEx, cCSAus, INMETRO, NEPSI, JPN, UK Ex
|
Product safety |
CE, C-TICK
|
Functional safety |
Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
|
Metrological approvals and certificates |
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)
|
Pressure approvals and certificates |
PED, CRN, AD 2000
|
Material certificates |
3.1 material
NACE MR0175/MR0103, PMI; welding test acc. to EN ISO, ASME, NORSOK
|
Reviews